Respironics Lifecare PLV-100

VENTILATOR

Product Description
Specifically designed for the long-term home use, the Respironics Lifecare PLV-100 is a reliable, easy to use portable volume ventilator. The PLV-100 is a versatile device that’s effective on a wide range of patients including pediatric patients and patients with minute ventilation demands.

FEATURES:
Wide Range Of Applications:
• Use in home, sub-acute care andchronic care settings for transport
• Small size and light weight ideal for mobile patients in wheelchairs, cars, and RVs
• Capable of ventilating pediatric to adult patients
• O2 bleed-in up to 40%
• Provides volume or pressure ventilation

Digital Displays:
• Tidal Volume
• I:E Ratio
• Respiratory Rate
• Inspiratory Flow Rate

Alarm Interface Capabilities:
• Remote Alarm
• Pressure Alarm
• Hospital/Nurse Call System

Three Modes Of Operation:
• Control l Assist-Control
• Spontaneous Intermittent Ventilation (SIMV)

Multiple Power Sources:
• 120 VAC/240 VAC
• 12 VDC external capabilities (up to 20 hours)
• Internal battery (1 hour fully charged)

SPECIFICATIONS
Front Panel Controls
• Mode Selection: Control, Assist/Control, SIMV
• Tidal Volume: 0.05 to 3.00 liters ± 10%
• Breath Rate: 2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2
• Inspiratory Flow Rate: 10 to 120 L/Min
• Battery Test Switch: Test Internal/External Battery
• Main Power Switch: On/Off, Recharge
• DC Circuit Breaker: Push/Pull Actuation
• AC Circuit Breaker: Push/Pull Actuation
• Airway Pressure Limit: 5 to 100 cm H2O ± 5 cm H2O
• Patient Airway Pressure: -10 to 100 cm H2O ± 5
• Sensitivity: +3 cm H2O ± 1 to at least -6 cm H2O
• Low Pressure Alarm: 2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5

Alarms
The audible alarm ranges between 70 and 85decibles (at 1 meter) and has its own power source separate from the internal battery. The power source and batteries are constantly recharged with a 2 amp trickle charge when the unit is plugged into AC power, whether it is operating or not.
• Low Pressure: Visual/Continuous Audible
• High Pressure: Intermittent Audible
• Apnea: Visual/Continuous Audible
• Inverse I:E Ratio: Visual Warning
• Increased Inspiratory Flow: Visual Warning
• Low Internal Battery: Visual/Continuous Audible
• Low External Battery: Visual/Continuous Audible
• Reverse Battery Connection: Continuous Audible
• Power Failure: Continuous Audible
• Power Source: Audible
• Unit Malfunction: Intermittent Audible