MODEL |
Infant Star 200 |
MODES OF OPERATION |
|
Assist/control |
Yes |
POWER NEEDED, VAC (ENERGY) |
108-132/207-242 |
SIZE, H x W x D, cm (in) (MODES OF OPERATION) |
22.9 x 39.4 x 25.4 (9 x 15.5 x 10) |
WEIGHT, kg (lb) (DISPLAY) |
9.1 (20) w/blender; 7.7 (17) w/out blndr |
REGULATORS |
|
Inspiratory time, sec (CONTROLS) |
0.1 |
Ventilation frequency, breaths/min (REGULATORS) |
1-150 |
Inspiratory flow, l/min (REGULATORS) |
0 |
Peak inspiratory pressure, cm H2O (REGULATORS) |
0-70 |
Cpap/peep, cm h2o (REGULATORS) |
0-20 |
MANUAL INSPIRATION (ALERTS) |
Yes |
TRANSPORTABLE (REGULATORS) |
Yes |
AUXILIARY POWER (REGULATORS) |
12VDC battery |
MONITORED PARAMETERS |
|
PIP, cm H2O (MONITORED PARAMETERS) |
Yes |
Map, cm h2o (PRESSURE ALERTS) |
Yes, 0-99.9 |
Cpap/peep, cm h2o (REGULATORS) |
Yes |
I:e ratio (REGULATORS) |
Yes |
Reverse I:E ratio (MONITORED PARAMETERS) |
Yes |
PATIENT ALERTS. O2 (ALERTS) |
|
Low PIP (PATIENT ALERTS. O2) |
Yes, 1-85 cm H2O |
High pip (REGULATORS) |
Yes, 1-85 cm H2O |
Low CPAP/PEEP (PATIENT ALERTS. O2) |
Yes |
IMV/SIMV |
Yes/yes |
ALERTS (CONFIGURATION) |
|
Fail to cycle (ALERTS) |
Yes |
Gas supply malfunction (ALERTS) |
Yes |
Power loss (ALERTS) |
Yes |
Vent inoperative (EQUIPMENT ALARMS) |
Yes |
Fio2 (PATIENT ALARMS) |
0.21-1 |
OTHER SPECIFICATIONS (AIR SUPPLY) Tracking relief valve; insufficient expiratory time alarm; maximum inverse I:E ratio alarm; apnea alarm; CPAP backup mode. |
|